Selamat Datang di Indotrade.co.id
Masuk | Daftar

Selamat Datang di

MIKASE LABORATORY - PT MITRA KATIGA SEJAHTERA / SUPPLIER ALAT UJI KUALITAS UDARA, AIR, CUACA , KEBISINGAN SERTA GETARAN

http://indotrade.co.id/mikaselaboratory/

Anggota Prioritas Prioritas Indotrade
Tahun ke-17

menampilkan 54 dari 233
  • Katalog Produk

    pH, conductivity, dissolved oxygen, and temperature, Model : HD3405.2 - HD3406.2 - HD3409.2 - HD3456.2, Brand : DeltaOhm

    [29 Nov. 2010, 23:17:27]
    pH,  conductivity,  dissolved oxygen,  and temperature,  Model : HD3405.2 - HD3406.2 - HD3409.2 - HD3456.2,  Brand : DeltaOhm
    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    Introduction
    The instrument series HD34… is made up of 4 bench top instruments for electrochemical measures: pH, conductivity,
    dissolved oxygen, and temperature.
    The displayed data can be stored ( datalogger) and can be transferred to PC or serial printer thanks to the multistandard serial ports RS232C and USB2.0 and software DeltaLog9 ( Vers.2.0 and subsequent ones) . The storing and printing parameters can be set from menu.
    The HD3405.2 measures pH, redox potential ( ORP) in mV.
    It measures temperature with Pt100 or Pt1000 immersion,
    penetration or contact probes. The pH electrode calibration can be carried out on one, two or three points and the calibration sequence can be chosen from a list of 13 buffers.
    The HD3406.2 measures conductivity, liquid resistivity in liquids, total dissolved solids ( TDS) and salinity using combined 4-ring and 2-ring conductivity/ temperature probes. Temperature is measured by Pt100 or Pt1000
    immersion, penetration or contact probes.
    The probe calibration can be performed automatically in one or more of the 147¼ S, 1413¼ S, 12880¼ S or 111800¼ S/ cm conductivity calibration solutions.
    The HD3409.2 measures the concentration ( in mg/ l) of dissolved Oxygen in liquids, the saturation index ( in% ) and the temperature using SICRAM combined probes of polarographic type with two or three electrodes and integrated temperature sensor. Temperature is measured by Pt100-SICRAM or direct 4 wire-immersion, penetration
    or contact probes.
    Thanks to an internal pressure sensor, the instruments automatically compensate for barometric pressure. The
    instrument anticipates automatic compensation of the Oxygen probe membrane permeability and of the salinity of the liquid being examined. The dissolved Oxygen probe’ s quick calibration function guarantees timely correctness of the performed measurements.
    The HD3456.2 measures pH, mV, redox potential ( ORP) , conductivity, resistivity in liquids, total dissolved solids ( TDS) , and salinity using combined 4-ring and 2-ring conductivity/ temperature probes. Temperature is
    measured by Pt100 or Pt1000 immersion, penetration or contact probes.
    The pH electrode calibration, as well as manual, can be carried out on one, two or three points and the calibration
    sequence can be chosen from a list of 13 buffers.
    The probe calibration can be performed automatically in one or more of the 147¼ S, 1413¼ S, 12880¼ S or 111800¼ S/ cm conductivity calibration solutions.
    The display shows continually the temperature in ° C or ° F and one selectable parameter according to the connected
    probe type, i.e. in case of conductivity probe it is possible to select between Ç or © or TDS or g/ l.
    Other common function of this instrument series include: Max, Min and Avg function, the Auto-HOLD function, the
    automatic turning off which can also be disabled.
    The instruments have IP66 protection degree.

    Technical characteristics of the instrument series HD34…
    Common technical data
    Instrument
    Dimensions ( Length x Width x Height) 220x120x55mm
    Weight 460g ( complete with batteries)
    Materials ABS, rubber
    Display 2x4½ characters plus symbols
    visible area: 52x42mm
    Operating conditions
    Working temperature -5 … 50° C
    Stocking temperature -25 … 65° C
    Working relative humidity 0 … 90% RH without condensation
    Protection degree IP66
    Power
    Batteries 3 batteries 1.5V type AA
    Autonomy ( only batteries) 100 hours with 1800mAh alkaline batteries
    Mains ( cod. SWD10) Output mains adapter 100-240Vac/ 12vdc-1A
    Security of memorized data
    Unlimited
    Selectable storage interval
    1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1hour
    Time
    Date and hour Schedule in real time
    Accuracy 1min/ month max departure
    Serial interface RS232C
    Type RS232C electrically isolated
    Baud rate Can be set from 1200 to 38400 baud
    Data bit 8
    Parity None
    Stop bit 1
    Flow Control Xon/ Xoff
    Serial cable length Max 15m
    Selectable print interval immediate or 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min and 1ora
    USB Interface
    Type 1.1 - 2.0 electrically isolated
    Common connections to all models
    Serial interface and USB 8-pole MiniDin connector
    Mains adapter ( cod. SWD10) 2-pole connector ( positive at centre) 12Vdc/ 1A
    EMC Standard regulations
    Security EN61000-4-2, EN61010-1 level 3
    Electrostatic discharge EN61000-4-2 level 3
    Electric fast transients EN61000-4-4 level 3, EN61000-4-5 level 3
    Voltage variations EN61000-4-11
    Electromagnetic interference
    susceptibility IEC1000-4-3
    Electromagnetic interference
    emission EN55020 class B

    Technical characteristics HD3405.2 pH - mV - ° C/ ° F measurement
    Measured values
    pH, mV, ° C, ° F
    Storage of measured values
    Type 2000 pages of 17 samples each
    Quantity 34, 000 sets of measures made up of [ pH or mV] or and [ ° C or ° F] .
    Measurement connections
    Temperature probe input 8-pole male DIN45326 connector with SICRAM module or TP47 module female BNC
    pH/ mV input
    pH Measurement
    Measurement range -2.000… + 19.999pH
    Resolution 0.01 or 0.001pH selectable from menu
    Accuracy ± 0.001pH ± 1digit
    Input impedance > 1012©
    Calibration error @ 25° C | Offset| > 20mV
    Slope 63mV/ pH
    Sensitivity 106.5%
    Automatic / manual -50… + 150° C
    temperature compensation
    mV Measurement
    Measurement range -1999.9… + 1999.9mV
    Resolution 0.1mV
    Accuracy ± 0.1mV ± 1digit
    Drift after 1 year 0.5mV/ year
    Temperature Measurement
    Pt100 measurement range -200… + 650° C
    Pt1000 measurement range -200… + 650° C
    Ni1000 measurement range -50… + 250° C
    Resolution 0.1° C
    Accuracy ± 0.1° C ± 1digit
    Drift after 1 year 0.1° C/ year
    Automatically detected pH standard solutions ( @ 25° C)
    1.679pH - 2.000pH - 4.000pH - 4.008pH - 4.010pH - 6.860pH - 6.865pH - 7.000pH - 7.413pH - 7.648pH - 9.180pH - 9.210pH - 10.010pH